M. Fero 11/05
Adapted from Turner's protocol (PDF) (website)
Fluorometric
Quantitation of RNA or DNA
Materials
DNA: Höchst 33258 (High range
solution):
(Invitrogen H1398
or Sigma B1155)
1µg/mL in
1x
TNE
Filter and store at
4ºC in an amber bottle.
10x TNE stock: 100 mM Tris, 2 M NaCl, 10 mM EDTA, pH7.4. Store
at room temp.
RNA: Ribogreen 200x
(Invitrogen R-11491)
detection reagent. Store dessicated at -20ºC.
Turner Designs TD-360 Fluorometer
DNA: Long UV LED
(underneath) and filter set (inside chamber)
RNA: Blue LED and filter set
RNA or DNA standards (100 ng/µL)
TE pH 8 (DNA), pH 7.5 (RNA)
Discussion
Fluorometry utilizes fluorescent dyes which specifically bind DNA or
RNA. It requires a negative control (to set the zero point on the
fluorometer) and a standard of known concentration.
The fluorometer shines light on the sample (excitation) and then
measues level of fluorescent light being emitted to the side (at a
90º angle) of the excitation light beam. The
fluorescent dyes are relatively specific to nucleic acids as opposed to
protein and other cellular components. The fluorescence of the
dyes increases when they bind nucleic acids. Fluorometry is about
1,000x more sensitive than spectrophotometric absorbance (i.e.
measurement of A260) and less susceptible to protein and
RNA contamination. However it also does not give a
crude measurement of purity (like an A260/A280
ratio) nor does it assure that the DNA or RNA is not degraded (e.g.
like size determination by gel electrophoesis). Do not use glass
(spectrophotmetry) cuvettes in a fluorometer because the frosted glass
on the side of the cuvette interferes with detection of fluorescent
light.
Detection Ranges
|
Large
Cuvette
|
Mini-cuvette
|
High Range Solution
|
20 ng - 1 µg
|
1 - 50 ng
|
Low Range Solution
|
1 - 50 ng
|
0.1 - 5 ng
|
Procedure (High range assay
solution)
- Insert the correct LED and filter set in the
fluorometer for either RNA or DNA.
- Turn on the Fluorometer to warm up for 5 min.
- DNA: Remove Höchst (high range) solution from the
refrigerator. Dilute Höchst 1/10 (v/v) in 1x TNE if a low
range assay is being performed.
RNA: Thaw Ribogreen and dilute to 1/200 (high range) or 1/2000 (low
range) in
1x RNAse-free TE (pH 7.5).
Make up enough reagent for the samples plus
a blank and at least one control.
- Large cuvettes require 1.5 mL/sample
- Mini-cuvettes need 100 µL/sample.
- Aliquot 1.5 mL (large cuvette) or 0.1 mL
(microcuvette) of either Höchst or Ribogreen reagent into 1.5 mL
eppendorf tubes.
- Add 1.5 µL of DNA or RNA to each tube and
vortex. In addition to samples this must also include a known
standard and a TE or H2O
control.
- Using narrow (gel loading) tips pipet the TE or H2O
control in a cuvette, close the chamber's lid and press the "Blank"
button. Press "1" when prompted to save the data. Open the
chamber's lid and transfer the control mix back to its original
Eppendorf tube.
- Transfer the DNA or RNA standard to the cuvette and
press the "Calibrate" button. When prompted enter the (undiluted)
concentration of the standard (i.e. 100) plus the "Enter" button. Press
"1" when prompted to save the data.
- Repeat the zero and calibration steps (6 and
7). This is important to avoid negative values at low
concentrations.
- Measure each RNA sample and record the
concentrations.
Note: The procedure for "Low range" assays is identical but it
uses the detection reagent diluted 1/10x the cocentrations listed above
to minimize background autofluorescence.
Fred Hutchinson Cancer Research Center
1100 Fairview Ave. N. PO Box 19024 Seattle, WA 98109
©2009 Fred Hutchinson Cancer Research Center, a nonprofit organization.
Terms of Use & Privacy Policy.